Base Strain Sensitivity Bar

During my internship at PCB Piezotronics, I designed and built a base strain sensitivity test station for Endevco’s Irvine facility. The station uses a steel bar and strain gauges to measure how bending an accelerometer’s mounting surface affects its output. This gave the Irvine team a way to test accelerometers in-house instead of shipping them to PCB’s facility in Depew, New York.

Theory

When the steel bar bends, the four strain gauges in the Wheatstone bridge change resistance, producing a voltage change that indicates the strain at the sensor mount. With 10 V excitation and an assumed gauge factor of 2.0, a 5 mV change in bridge output indicated 250 microstrain, and the base strain response was determined by subtracting the Unit Under Test’s unloaded output voltage from its output voltage at 250 microstrain.


CAD and FEA

  • Modeled the steel cantilever bar in Autodesk Inventor and designed mounting hole patterns to accommodate multiple accelerometer models.

  • Used finite element analysis (FEA) to predict bar deflection and surface strain at each sensor mount, then updated the model with measured bar dimensions after machining.


Vise and Bar Installation

  • Drilled mounting holes in the concrete lab floor with a rotary hammer and secured the machine vise with concrete anchors.

  • Clamped the steel bar vertically in the vise to form a cantilever, leaving the upper end free to bend during base strain testing.


Strain Gauge Installation and Wiring Arrangement

Back

Data Acquisition and Testing

Front

Bridge Diagram

  • The steel surface was abraded, cleaned, and conditioned before the gauge locations were marked for alignment.

  • Four strain gauges were aligned with the marks and bonded to the bar, two per face. The leads were soldered to form a full Wheatstone bridge, and the gauges and connections were covered with Kapton tape for protection.

  • As the bar bent, the gauges on one face increased in resistance under tension, while those on the opposite face decreased in resistance under compression. The bridge combined these changes into a voltage signal proportional to bending strain.


Controller and Multimeter Setup

Sample Sensor

Sample Sensor

  • Five accelerometers were tested on the strain bar using screw and stud mounts. The screw-mounted units were tested in four orientations, with three push-pull trials per orientation.

  • The controller routed each signal to a separate multimeter. Multimeter 2 monitored the strain-gauge bridge to identify the 250-microstrain target, while Multimeter 1 measured the Unit Under Test’s output voltage for comparison with its unloaded baseline.